Handheld digital manometer PG-100N
Handheld digital manometer PG-100N
- 
Excellent stability
High precision semiconductor chip combined with temperature compensation circuit guarantee excellent stability.. - 
Long battery life
Lasts more than 1000hrs with two AAA dry batteries. - 
Automatic power off function
 - 
Easy, one-handed operation
 - 
Pressure unit selectable type also available (PG-100N-****-W).
 
Model number designation
| PG-100N- | 103R | M | X2 | T | 
|---|---|---|---|---|
| 
 Series name  | 
 Rated pressure range/Pressure reference 101R: -10~10 kPa/ Gauge(Compound) 102R: -100~100 kPa/ Gauge(Compound) 103R: -100~1000kPa Gauge(Compound)  | 
 Unit of display Blank: kPa M: MPa H: mmHg W: Selectable✽1  | 
 Inspection report/Calibration Certificate Blank: Not attached X1: Inspection report X2: Inspection report and Calibration certificate  | 
 Traceability system chart Blank: Not attached T: Traceability system chart  | 
(Refer to the Pressure unit selectable W type below.)
Pressure unit selectable W type
| Pressure range | Unit selectable W type | ||||||
|---|---|---|---|---|---|---|---|
| kPa | Mpa | mmHg | in.Hg | mmAq | psi | bar | |
| 101R | 〇 | ― | 〇 | 〇 | 〇 | 〇 | 〇 | 
| 102R | 〇 | 〇 | 〇 | 〇 | ― | 〇 | 〇 | 
| 103R | 〇 | 〇 | ― | 〇 | ― | 〇 | 〇 | 
※ " ー" is not supported as a display unit.
List of model numbers
| Pressure reference | Gauge(Compound) | |||
|---|---|---|---|---|
| Rated pressure range (kPa) | - 10 ~ 10 | - 100 ~ 100 | -100 ~ 1000 | |
| kPa | Product only | PG-100N-101R | PG-100N-102R | PG-100N-103R | 
| Inspection report | PG-100N-101R-X1 | PG-100N-102R-X1 | PG-100N-103R-X1 | |
| Inspection report and Traceability system chart | PG-100N-101R-X1T | PG-100N-102R-X1T | PG-100N-103R-X1T | |
| Inspection report and Calibration certificate | PG-100N-101R-X2 | PG-100N-102R-X2 | PG-100N-103R-X2 | |
| Traceability system chart | PG-100N-101R-T | PG-100N-102R-T | PG-100N-103R-T | |
| Inspection report,Calibration certificate and Traceability system chart | PG-100N-101R-X2T | PG-100N-102R-X2T | PG-100N-103R-X2T | |
| MPa | Product only | ―――― | ―――― | PG-100N-103R-M | 
| Inspection report | PG-100N-103R-M-X1 | |||
| Inspection report and Traceability system chart | PG-100N-103R-M-X1T | |||
| Inspection report and Calibration certificate | PG-100N-103R-M-X2 | |||
| Traceability system chart | PG-100N-103R-M-T | |||
| Inspection report,Calibration certificate and Traceability system chart | PG-100N-103R-M-X2T | |||
| mmHg | Product only | PG-100N-101R-H | PG-100N-102R-H | ―――― | 
| Inspection report | PG-100N-101R-H-X1 | PG-100N-102R-H-X1 | ||
| Inspection report and Traceability system chart | PG-100N-101R-H-X1T | PG-100N-102R-H-X1T | ||
| Inspection report and Calibration certificate | PG-100N-101R-H-X2 | PG-100N-102R-H-X2 | ||
| Traceability system chart | PG-100N-101R-H-T | PG-100N-102R-H-T | ||
| Inspection report,Calibration certificate and Traceability system chart | PG-100N-101R-H-X2T | PG-100N-102R-H-X2T | ||
| Selectable | Product only | PG-100N-101R-W | PG-100N-102R-W | PG-100N-103R-W | 
| Inspection report | PG-100N-101R-W-X1 | PG-100N-102R-W-X1 | PG-100N-103R-W-X1 | |
| Inspection report and Traceability system chart | PG-100N-101R-W-X1T | PG-100N-102R-W-X1T | PG-100N-103R-W-X1T | |
| Inspection report and Calibration certificate | PG-100N-101R-W-X2 | PG-100N-102R-W-X2 | PG-100N-103R-W-X2 | |
| Traceability system chart | PG-100N-101R-W-T | PG-100N-102R-W-T | PG-100N-103R-W-T | |
| Inspection report,Calibration certificate and Traceability system chart | PG-100N-101R-W-X2T | PG-100N-102R-W-X2T | PG-100N-103R-W-X2T | |
Standard specifications [kPa version]
◆Unless otherwise specified, the specs are defined at an ambient temperature of 25 ± 5 °C.
| Item | PG-100N | ||||
|---|---|---|---|---|---|
| 
 General  | Pressure reference | Gauge | |||
| Model number | 101R | 102R | 103R | ||
| Rated pressure range | - 10 ~ 10 kPa | - 100 ~ 100 kPa | - 100 ~ 1000 kPa | ||
| Maximum pressure | 20 kPa | 200 kPa | 1500 kPa | ||
| Break-down pressure | 50 kPa | 500 kPa | 2000 kPa | ||
| Pressure medium | Non-corrosive gases | ||||
| Operating temp. range | 0 ~ 50°C | ||||
| Compensated temp. range | 0 ~ 50 °C | ||||
| Operating humidity | 15 ~ 85 %RH(No condensation) | ||||
| Storage temp. | 
 - 20 ~ 70 °C(Atmospheric pressure, humidity 65 %RH maximum)  | 
||||
| Pressure port | M5 female | ||||
| Net weight | Approx. 100g | ||||
| Power | AAA dry battery 1.5 VDC × 2 Battery life 1000 h minimum | ||||
| 
 Display  | No. of digits | 3 1/2 | |||
| Display element | LCD | ||||
| Rated display range | - 10.00 ~ 10.00 kPa | - 100.0 ~ 100.0 kPa | - 100 ~ 1000 kPa / -. 100 ~ 1.000(MPa version) | ||
| Display cycle | Approx. 4 times / s | ||||
| Integrated accuracy | ± 0.5 %F.S. ± 2 digits (at 25 °C ± 5 °C) | ||||
| ± 1.5 %F.S. ± 2 digits (at 25 °C ± 25 °C)  | 
± 1.0 %F.S. ± 2 digits (at 25 °C ± 25 °C)  | 
||||
| Resolution | 1 digit | ||||
| 
 Analog   | Power supply (External supply)  | 
5 ± 0.5 VDC 5 mA maximum | |||
| Output voltage | 0 ~ ± 1.000 V ※Proportionate to display value | ||||
| Output accuracy/Resolution | ± 10 mV / 1mV | ||||
| Output resistance | Approx. 10 k Ω | ||||
| Response | Approx. 10 ms | ||||
| Additional function | Auto power off, Display hold, Battery replacement display, (LOBAT), Zero point adjustment | ||||
| Accessories | 2 fittings (half union for 6mm O.D. tube, barb fitting for 4mm I.D. tube), 2 AAA batteries, strap | ||||
| Option  (Sold seperately)  | 
PG-100N soft carrying case | ||||
Standard specifications [Non SI unit version]
◆Unless otherwise specified, the specs are defined at an ambient temperature of 25±5 °C.
| Item | PG-100N | |||
|---|---|---|---|---|
| 
 General  | Pressure reference | Gauge | ||
| Model number (mmHg display : compound)  | 
101R-H | 102R-H | ||
| Rated pressure range kPa(mmHg)  | 
- 10 ~ 10 (- 75.0 ~ 75.0) | - 100 ~ 100 (- 750 ~ 750) | ||
| Maximum pressure kPa(mmHg)  | 
20 kPa (150) | 200 (1500) | ||
| Break-down pressure kPa(mmHg)  | 
50 kPa (375) | 500 (3750) | ||
| Pressure medium | Non-corrosive gases | |||
| Operating temp. range | 0 ~ 50 °C | |||
| Compensated temp. range | 0 ~ 50 °C | |||
| Operating humidity | 15 ~ 85 %RH(No condensation) | |||
| Storage temp. | - 20 ~ 70 °C(Atmospheric pressure, humidity 65 %RH maximum) | |||
| Pressure port | M5 female | |||
| Net weight | Approx. 100g | |||
| Power | AAA dry battery 1.5 VDC × 2 Battery life 1000 h minimum | |||
|  
 Display  | No. of digits | 3 1/2 | ||
| Display element | LCD | |||
| Rated display range (mmHg display)  | 
- 75.0 ~ 75.0 | - 750 ~ 750 | ||
| Display cycle | Approx. 4 times/s | |||
| Integrated accuracy | ± 0.5 %F.S. ± 2 digits (at 25 °C ± 5 °C) | |||
| ± 1.5 %F.S. ± 2 digits  (at 25 °C ± 25 °C)  | 
± 1.0 %F.S. ± 2 digits  (at 25 °C ± 25 °C)  | 
|||
| Resolution | 1 digit | |||
| 
 Analog   | Power supply (External supply)  | 
5 ± 0.5 VDC 5 mA maximum | ||
| Output voltage | 0 ~ ± 1.000 V ※Proportionate to display value | |||
| Output accurac/Pesolution | ± 10 mV / 1mV | |||
| Output resistance | Approx. 10 kΩ | |||
| Response | Approx. 10 ms | |||
| Additional function | Auto power off , Display hold,  Battery replacement display(LOBAT), Zero point adjustment  | 
|||
| Accessories | 2 fittings (half union for 6mm O.D. tube,  barb fitting for 4mm I.D. tube), 2 AAA batteries, strap  | 
|||
| Option (Sold seperately)  | 
PG-100N soft carrying case | |||
Environmental characteristics
| Test item | Test conditions(At 25 ± 5 °C) | Permissible change | 
|---|---|---|
| Vibration | Amplitude 1.5 mm P-P, 10 ~ 55 Hz,  3 directions for 2 hours each (packed)  | 
Indicating accuracy : ± 1 %F.S. maximum | 
| Shock | 98.1 m/s2 , 3 directions for 3 times each (packed) | |
| Moisture resistance | 40 °C, 90 ~ 95 %RH, 240 hrs | |
| Pressure cycling | 0 ~ Rated pressure, 106 cycles | 
Outline dimensions
Unless otherwise specified, tolerance : ± 0.5 (Unit: mm)

Troubleshooting
| Symptom/Problem | Possible cause | Action/Troubleshooting | |
|---|---|---|---|
| Fluctuation in display value. | Fluctuation in display value. | Fluctuation in measured pressure or leakage. | 
  | 
| Zero point deviation | Zero point deviation | Residual pressure in the pipework. | 
  | 
| No display when power is on. | No display when power is on. | Reversed polarity of dry batteries. | 
  | 
| Display shows "LOBAT". | Display shows "LOBAT". | Low battery. | 
  | 
| Flashing display. | Flashing display. | A pressure exceeding 130% of the rating is applied. | 
  | 